qc-plate800

QC-plate 800


The QC-plate 800 checks the mechanical alignment of both the plate transport and the optical path and tests
the accuracy, linearity and precision of single- and multichannel microplate reader.

The QC-plate consists of a precise milled aluminum body in the dimensions of a SBS standard microplate.
In positions A1 to H1 as well as A12 to H12 there are 12 precise drilled holes to check the plate alignment.
Neutral density filters in columns 3, 5, 7 and 9 with nominal absorbance of 0.2; 1.2; 2.0 and 2.5 are used to verify the accuracy, precision and linearity.

Each QC-plate 800 is supplied with a calibration sheet showing the absorbance for each neutral density filter at different wavelengths. The calibration data can be traced back to NIST standard.

A QC-program is available for several microplate readers which automates the whole QC test. Please contact us to see if your reader is supported.


Copyright © 2013 - MPI-Design

All Rights Reserved